| SYSTEM
Design:
F/number:
Focal length:
Grating size:
Main mirrors:
Focal plane:
Entrance slit:
|
two spectrographs in one housing,
vertically-symmetric configuration
9,3
158 mm
15 x 15 x 3 mm
spherical
30 x 0,6 mm
0,025 x 0,6 mm
|
IMAGING
Imaging spatial resolution (aberration limited):
|
~30 µm
|
WAVELENGTH
Grating, l/mm:
Blazing wavelength, nm:
Reciprocal dispersion (average), nm/mm:
Wavelength resolution, nm:
Detectable region, nm:
Wavelength range where grating
efficiency is more than 40%, nm:
|
400
700
49,3
<0,3
450
470-1100 |
600
450
33,5
<0,2
310
300-900 |
1200
280
17,7
<0,1
150
190-560 |
1800
270
11,8
<0,07
100
190-540 |
Echelle, 75 l/mm
-
0,3 (200 nm, 120 order)
<0,006
-
190-1100 |
Each of the specified gratings or other grating can be installed in any of two Spectrographs (channels) |
DETECTION SYSTEM
Linear sensor:
Number of pixels:
Width of pixels, µm:
Height of pixels, mm:
Spectral response, nm:
Min. exposure time, ms:
Max. exposure time at 20°C, s:
Min. time between scans, µs:
Dynamic range for 1 scan:
Analog to digital converter:
Synchronization:
Computer interface:
|
S8378-1024Q
1024
25
0,5
200-1060
2,1
5,0
1,0
3600:1
14 bit, 16384 counts
internal, external
USB
|
TCD 1304AP
3648
8
0,2
190-1100
7,4
0,5
1,0
1100:1
12 bit, 4096 counts
internal, external
USB
|
|
FIBER OPTICS
Optical input:
Fiber optical cable:
Material:
Length: |
direct via a condenser or fiber-optical cable
two channels (one common input, two outputs)
one channel (2 pcs: one input, one output)
UV silica; 0,4 mm dia.
0,5 m; 1 m; 2 m (other lengths available)
|
| Overall Size (LxWxH): 200 x 143 x 72 mm; Weight: 2,5 kg |
SOFTWARE “PSI-Line” (real time 32 bit, Win’9x/2000/XP):
Spectrometer channels mode: One / Two
Background measurement
Intensity calibration
WL calibration
Scan modes: Single spectrum / Multi spectra
Cyclic (scan) mode: NoneStop / Counter
Integration in WL range |
Accumulation: Summing / Averaging
Process of peaks: Find / Measure / Correlation / Report
Smooth spectrum
Photometer mode: Transmission / Absorption
Mathematical processor
Lambda-meter
Data compatibility with MS Office EXCEL |
|