Analytical Instruments
Spectral Instruments
Lasers
Optics
 
  SPECTRAL INSTRUMENTS  
 
    Home   About Us   Contact Us
 
Double Dispersion Monochromator / Spectrograph MSDD1000
 

Double Dispersion
Monochromator / Spectrograph

model
MSDD1000

Specifications  |  Dimensional Drawing  |  Gratings  |  Accessories

MSDD1000 Double Dispersion Monochromator / Spectrograph is designed around the modified Cherny-Turner configuration with the two-level arrangement of optical components to allow dispersion doubling.
With a relatively small size, the MSDD1000 Monochromator / Spectrograph ensures the high resolution typical of instruments with a focal length of 1000 mm.

    Double Dispersion Monochromator / Spectrograph MSDD1000 - Features

  • Complete computer control
  • Compromise between small size and high
    resolution due to original design of modified
    Cherny-Turner double dispersion configuration
  • Two-grating pairs turret with automatic
    grating switching or single-grating pair
    holder for easy manual grating interchange
  • Abberation-free corrected optical system
  • Low stray light and no re-entrant spectra
  • Continuously automatically or manually
    variable slits
  • Two entrance ports with automatic port
    switching enable simultaneous mounting
    of different light sources
  • Two flat field output ports with automatic port
    switching allow simultaneous mounting
    of different detectors
  • Original software Win’9x/ME/2000/XP

    Double Dispersion Monochromator / Spectrograph MSDD1000 - Applications

  • UV to IR measurement
  • Emission, Fluorescence and Raman Spectroscopy
  • Absorption and Transmission
  • Multi-track Spectroscopy (CCD and CID)

Double Dispersion Monochromator / Spectrograph MSDD1000 - Optic Layout
Double Dispersion Monochromator / Spectrograph MSDD1000 - Optic Layout

 

Double Dispersion Monochromator / Spectrograph MSDD1000 - Spectrum


Specifications

SYSTEM
  Design:

  Ports:
  Usable wavelength range:
  F/Number:
  Focal length:
  Grating size:
  Grating mounts:

  Scanning range limited by the grating rotation angle:
  Main mirrors:
  Stray light: 
  Flat field:

Original design of modified Cherny-Turner two-fold
configuration with one or two pairs of gratings
2 input (axial + lateral), 2 output (axial + lateral)
180 nm to 60 µm
5,9
1000 mm (2 x 500 mm)
80 x 70 x 10 mm
- Automatic turret with two pairs of gratings
- Holder with a single pair of gratings
0-1280 nm (grating 1200 l/mm)
spherical
10-6 (20 nm from 632,8 laser line)
28 x 10 mm
DRIVE
  Motor:
  Drive:
  Step size:
  Precision:
  Max. speed:
high torque microstepper
sine mechanism
0,18 arc seconds
± 1 step
10000 steps/s
WAVELENGTH (grating 1200 l/mm)
  Reciprocal dispersion:
  Wavelength repeatability:
  Wavelength resolution:
  Wavelength accuracy:
  Wavelength step size (aver):
0,78 nm/mm
± 0,005 nm
0,012 nm
± 0,05 nm
0,0015 nm
SLITS
  Motor (micrometer) driven slit assembly:

  Step size:
  Parallelism:
  Accuracy (with 1 mm width):
  Reading accuracy (micrometer):
Width variable from 0 to 2,0 mm
Diaphragm-varied height from 0 to 10 mm
0,5 µm
± 1 µm
± 10 µm
2 µm
SYSTEM CONTROL
  Central control by on board CPU
  External control by Hand Controller
  Programmed control by external computer
  External interface RS-232
Overall Size (LxWxH): 556 x 324 x 340 mm;  Weight: 35 kg
Dimensional Drawing
Double Dispersion Monochromator / Spectrograph MSDD1000 - Dimensional Drawing

 
  Top | Home | About Us | Contact Us | Analytical Instruments | Spectral Instruments | Lasers | Optics